Add to Cart
LE100 Liquid Particle Counter Injections Testing In Pharma
Liquid Particle Counter for USP 788 sizes and counts suspended particulate matter in a wide range of liquids, including parenterals, to meet all current U.S., European and Japanese Pharmacopoeia standards. When it comes to particle monitoring quality control for liquid drug manufacturers, it’s important to meet regulations while minimizing waste.
| Technical parameters | liquid particle counter |
| Particle size range | 2-400 μm |
| Particle size channel | 24 |
| Maximum detection concentration | 10000 particles/ml |
| Accuracy | ±10% |
| Repeatability | RSD<2% (particles>1000 per ml, 5ml sample) |
| Resolution | <10% (U.S.P, ISO 21501) |
| Syringe specifications | 1ml, 5ml, 10ml*, 25ml |
| Sampling speed | 20 ml/min* |
| Sample volume accuracy | ±1% |
| Minimum sampling volume | 0.2ml** |
| Sensor material | 316 stainless steel, quartz glass, PTFE, NBR |
| Power supply | AC 100~240V, 50/60HZ |
| Mixing method | Magnetic stirrer (speed adjustable) |
| Data output | RS232 port; USB; built-in thermal printer |
| Display output | 7 inch LCD color touch screen |
| Size | 320(L)×309(W)×394(H) mm |
| Weight | 11.0 kg |
| Operating environment | Temperature 10-40 ℃; RH 0 ~ 95% non-condensing |
| Software system | LE-Assistant (optional) |
|
* Standard configuration ** Under 1ml syringe specifications the minimum sampling volume is 0.2ml, under the standard configuration, the minimum sampling volume is 0.4ml | |

| 8.Exhaust fan | 1. Display |
| 9.heat emission hole | 2. Door of sample chamber |
| 10. Power switch | 3. Thermal printer |
| 11. AC power outlet and fuse | 4. Syringe |
| 12. R232 port | 5. Sample tube |
| 13. Sample drain | 6. LED chamber light |
| 14. USB port | 7. Stirrer |
Components of liquid particle counter
| The instrument ×1 | Power cord ×1 |
| Fuse ×1 | Detecting cup ×1 |
| Stir rod ×1 | Operation manual ×1 |
| Drain line tube ×1 | Testing report ×1 |
| Touch pen ×1 | Quality certificate ×1 |
| Printing paper ×1 | Warranty card ×1 |
| Packing list ×1 |
Features of liquid particle counter
Built-in USP testing standards Precision syringe sampling device, accurate sampling volume, stable speed Built-in calibration curve, in line with calibration requirements of ISO-21501 Intelligent automatic detection and automatic calibration function, instrument performance judgment and calibration Use of magnetic stirring, stirring speed adjustable USB interface direct export data, paperless electronic print report Powerful application software development support. Easy to connect to the laboratory information management system etc. The LE-Assistant software implements PC control and provides professional data processing Compliant with FDA 21 CFR part 11 "Electronic Records and Signatures" and user rating management
Application of liquid particle counter
Injections testing under CP, USP, EP, JP and other standards Cleanliness detecting of medical equipment / parts Filtration efficiency detection Laboratory water testing 